NovoCyte 3000 Flow Cytometer

The NovoCyte 3000 is a high-throughput flow cytometric analyzer with three lasers (Violet 405 nm, Blue 488 nm, Red 642 nm) and filters for detecting up to 13 fluorescent colors. The instrument is equipped with the NovoSampler Pro multi-sampler, which accommodates 24-, 48-, or 96-well plates or a 24-tube rack. The Novocyte uses the NovoExpress software for data acquisition and analysis. Data files can be exported as FCS 2.0 or FCS 3.1.

Flow cytometry can measure the number of cells, the percentage of live cells, and the presence of intra- and extra-cellular markers.  A few common applications are cell cycle analysis, apoptosis assays, and phenotyping of mixed cell populations.

When planning a flow cytometry experiment and choosing fluorochromes, it is important to know the fluorochromes' specific excitation and emission wavelengths, as well as the lasers, filters, and detectors available on the instrument. Below is the information for the NovoCyte 3000.

NovoCyte 3000 Configuration and Fluorochromes
Excitation LaserChannelBandpass FilterDefault AliasExamples of Fluorochromes
Violet 405 nmV445445/45Pacific BlueDAPI, BV421, eFluor 450
V530530/30AmCyanBV510, cFluorV547, AF430, eFluor506
V572572/28Pacific OrangeQDot565*, BV570
V615615/20Qdot 605BV605, eFluor 605NC
V675675/30Qdot 655BV650, eFluor 700
V780780/60Qdot 800BV785, Qdot 800*
Blue 488 nm  B530530/30FITCFITC, AF488, CFSE, GFP, YFP, BB515
B572572/28PEPE, Cy3, eFluor 570, SYTOX Orange
B615615/20PE-Texas RedPI, PE-Texas Red, RFP
B675675/30PerCPBB660*, PE-Cy5, PI, 7AAD, PerCP-Cy5.5* PE-AF647
B780780/60PE-Cy7PE-Cy7, PE-AF750
Red 640 nmR675675/30APCAPC, AF647, AF660, eFluor660, DRAQ 5
R780780/60APC-Cy7APC-Cy7, APC-AF750, APC-H7, APC-eFluor780

*It is important to remember that many fluorescent dyes are excited by more than one laser and fluoresce in multiple channels. It is very helpful to use a spectral viewer when planning an experiment.