Fibrinogen Degradation Products (FDP)
Test Name: | Fibrinogen Degradation Products (FDP) |
Epic Order Code: | LAB761 |
CPT Code: | 85362 |
Specimen(s) Type: | Plasma; Full tube required. |
Acceptable Container(s): | Top is light blue. |
Testing Schedule: | 24 hours/day, 7 days/week |
Turn Around Time: | STAT: 1 hour |
Routine: 2 hours | |
Collection Information: | No special patient preparation required. Obtain specimen by standard collection procedures. If only FDP is being collected, draw 5 mL into a red top tube, discard and then collect FDP tube. A full tube is required. |
Transport Information: | Deliver to lab immediately. All specimens must be signed into the laboratory. |
Reference Clients: | If unable to deliver to the lab within 1 hour, centrifuge the specimen for 10 minutes at 3500 RPM to ensure platelet poor plasma. Remove plasma from cells, taking care to avoid the WBC/platelet buffy layer and place into a transport tube. Store and transport at room temperature for up to 4 hours. If testing is delayed beyond this, the sample must be stored and transported frozen at -20° C and delivered to lab within 2 weeks. |
Causes for Rejection: | Improperly labeled, incorrect container, contaminated, insufficient quantity, incorrect/delay in transport, grossly hemolyzed, clotted, under or over-filled tube. |
Reference Range: | < 5 µg/mL |
Additional Information: | Falsely elevated FDP levels may occur with elevated Rheumatoid Factor. |